FEATURES:
Digit as well Bar graphic display.
User selectable audiovisual alarm limits.
With active PIP releif valve.
Accurate & reliable.
Auto calibration.
Very fast reading.
Operated on main supply.
8 hrs Battery backup(optional).
Both upper & lower alarm limits selectable by only two key.
Size in cm 17 x 6 x 13.
Intended to use.
Technical Specification:
Ambient operating temperature range: |
10°C to 50°C. |
Operating Altitude: |
-557 ft. to +16,000 ft / -170m to +4877m. |
Patient pressure Range: |
0 to 100 cmH2O. |
Sample Pressure inlet port size: |
5 mm male connector. |
Transport/Storage Temperature: |
-13°F to +131°F (-25°C to +55°C). |
Accuracy: |
+/- 1 cmH2O. |
Humidity: |
15% to 95% non-condensing. |
Dimensions: |
(17cm x 6cm x 13cm). |
Weight: |
0.85 Kg. |
Power: |
230V AC (10W) |
Equipment Classification: |
Class 2 Device, Continuous Operation |
Water protection: |
Not protected |
Alarm: |
Audio Visual |
Alarm range: |
default low = 0, High = 35.
User selectable low = 0 TO 20.
User selectable High = 5 TO 90.
|
Display: |
LCD. |
Battery Backup (Optional): |
8 Hrs |